WaveAce 1000/2000 Remote Control Command Reference
WaveAce Remote Control Data Parameters When a command uses additional data values, they are expressed as ASCII data that can take the form of charact
Operator's Manual Multiplier Exponential Notation Suffix P 1E-12 pico- A 1E-18 atto- String Data This is used to send multiple characters as a s
WaveAce Remote Control The oscilloscope might respond with: TIME_DIV 50 NS;C1:COUPLING D50 Note the response message refers only to the two queries t
Operator's Manual Read ESR using the *ESR? query. The response is the binary weighted sum of the register bits. The register is cleared with *ESR
WaveAce Remote Control Commands and Queries by Subsystem Acquisition - Controlling Waveform Captures Short Form Long Form What the Command/Query Does
Operator's Manual Display - Displaying Waveforms Short Form Long Form What the Command/Query Does DTJN DOT_JOIN Controls the interpolation lines
WaveAce Remote Control Status - Obtaining Status Information Using Status Registers Short Form Long Form What the Command/Query Does *CLS *CLS Clears
Operator's Manual Commands and Queries by Name (Alphabetical) Short Form Long Form What the Command/Query Does *CAL? *CAL? Performs a complete in
WaveAce Remote Control Short Form Long Form What the Command/Query Does INTS INTENSITY Controls the brightness of the grid. OFST OFFSET Allows output
Operator's Manual Part II: Command Reference All remote control commands and queries recognized by the instrument can be executed in either a loc
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com WaveAce 1000/2000 Remote Contr
WaveAce Remote Control ACQUISITION - *TRG The *TRG command executes an ARM command. *TRG is the equivalent of the 488.1 GET (Group Execute Trigger) m
Operator's Manual ACQUISITION – ACQUIRE_WAY, ACQW The ACQUIRE_WAY command specifies the acquisition mode. The ACQUIRE_WAY? query returns the cur
WaveAce Remote Control ACQUISITION - ATTENUATION, ATTN The ATTENUATION command selects the vertical attenuation factor of the probe. Values up to 100
Operator's Manual ACQUISITION - BANDWIDTH_LIMIT, BWL The BWL command enables or disables the bandwidth limit (low-pass filter) of the specified i
WaveAce Remote Control ACQUISITION - COUPLING, CPL The COUPLING command selects the coupling mode of the specified input channel. The COUPLING? query
Operator's Manual ACQUISITION - INTERLEAVED, ILVD The INTERLEAVED command enables or disables Equivalent Time Mode for timebase settings where bo
WaveAce Remote Control ACQUISITION - OFFSET, OFST The OFFSET command allows adjustment of the vertical offset of the specified input channel. The OF
Operator's Manual ACQUISITION – PEAK DETECT, PDET The PEAK_DETECT command switches ON or OFF the peak detector built into the acquisition system.
WaveAce Remote Control ACQUISITION - TIME_DIV, TDIV The TIME_DIV command modifies the timebase setting. The new timebase setting can be specified wit
Operator's Manual ACQUISITION - TRIG_COUPLING, TRCP The TRIG_COUPLING command sets the coupling mode of the specified trigger source. The TRIG_CO
Operator's Manual TABLE OF CONTENTS Part I: Introduction to Remote Control ...
WaveAce Remote Control ACQUISITION - TRIG_DELAY, TRDL The TRIG_DELAY command sets the time at which the trigger is to occur with respect to the nomin
Operator's Manual ACQUISITION - TRIG_LEVEL, TRLV The TRIG_LEVEL command adjusts the trigger level of the specified trigger source. An out-of-rang
WaveAce Remote Control ACQUISITION - TRIG_MODE, TRMD The TRIG_MODE command specifies the trigger mode. The TRIG_MODE? query returns the current trigg
Operator's Manual ACQUISITION - TRIG_SELECT, TRSE The TRIG_SELECT command selects the condition that triggers the acquisition of waveforms. Only
WaveAce Remote Control ACQUISITION - TRIG_SLOPE, TRSL The TRIG_SLOPE command sets the trigger slope of the specified trigger source. The TRIG_SLOPE?
Operator's Manual ACQUISITION - VOLT_DIV, VDIV The VOLT_DIV command sets the vertical sensitivity in Volts/div. The probe attenuation factor is
WaveAce Remote Control ACQUISITION - WAIT The WAIT command prevents your instrument from analyzing new commands until the current acquisition has bee
Operator's Manual CURSOR - CURSOR_MEASURE, CRMS The CURSOR_MEASURE command specifies the type of cursor or parameter measurement to be displayed,
WaveAce Remote Control CURSOR - CURSOR_SET, CRST The CURSOR_SET command allows you to position any one of the independent cursors at a given grid loc
Operator's Manual CURSOR - CURSOR_VALUE?, CRVA? The CURSOR_VALUE? query returns the values measured by the specified cursors. The PARAMETER_VALU
WaveAce Remote Control DISPLAY - HOR_MAGNIFY, HMAG ...
WaveAce Remote Control CURSOR - PARAMETER_CUSTOM, PACU The PArameter_CUstom command controls the placement of custom cursors and can also be used to
Operator's Manual Parameter Definition MAX Maximum value MEAN Mean value MIN Minimum value NDUTY - Duty cycle NWID Negative width (negative edge
WaveAce Remote Control CURSOR - PARAMETER_VALUE?, PAVA? The PArameter_VAlue query returns the current values of the pulse waveform parameters and mas
Operator's Manual Parameter Definition MAX Maximum value MEAN Mean value MIN Minimum value NDUTY - Duty cycle NWID Negative width (negative edge
WaveAce Remote Control DISPLAY - DOT_JOIN, DTJN The DOT_JOIN command controls the interpolation lines between data points. Setting DOT_JOIN ON selec
Operator's Manual DISPLAY - HOR_MAGNIFY, HMAG The HOR_MAGNIFY command horizontally expands traces (zooms) by a specified factor. Magnification fa
WaveAce Remote Control DISPLAY - HOR_POSITION, HPOS The HOR_POSITION command repositions the geometric center of a zoom to display different portions
Operator's Manual DISPLAY - INTENSITY, INTS The INTenSity command sets the intensity level of the grid and trace. Command Syntax INTenSity GRID,&
WaveAce Remote Control DISPLAY - PERSIST_SETUP, PESU The PERSIST_SETUP command sets the duration of the display, in seconds, when in persistence mode
Operator's Manual DISPLAY – SCREEN SAVE, SCSV The SCREEN_SAVE command controls the automatic Screen Saver, which automatically shuts down the int
Operator's Manual Part I: Introduction to Remote Control About This Manual This manual includes a complete list of the command you’ll need to per
WaveAce Remote Control DISPLAY - TRACE, TRA The TRACE command enables or disables the display of a trace. An environment error is set if an attempt i
Operator's Manual MISCELLANEOUS - *CAL? The *CAL? query causes the oscilloscope to perform an internal self-calibration and generates a response
WaveAce Remote Control MISCELLANEOUS – AUTO-CALIBRATE, ACAL Available only on WaveAce 2000 models, AUTO_CALIBRATE enables or disables automatic calib
Operator's Manual MISCELLANEOUS - COMM_HEADER, CHDR The COMM_HEADER command controls the way the oscilloscope formats responses to queries. There
WaveAce Remote Control MISCELLANEOUS - DEFINE, DEF The DEFINE command specifies a mathematical function to be performed. This command is used to cont
Operator's Manual MISCELLANEOUS - DELETE_FILE, DELF The DELETE_FILE command deletes a file from the currently selected directory. Command Syntax
WaveAce Remote Control MISCELLANEOUS – FILENAME, FLNM The FILENAME command is used to change the default filename given to any traces, setups or hard
Operator's Manual SAVE/RECALL SETUP - *RCL The *RCL command sets the state of your instrument, using one of the six non-volatile panel setups (Pa
WaveAce Remote Control SAVE/RECALL SETUP - *SAV The *SAV command stores the current state of your instrument in non-volatile internal memory. The *SA
Operator's Manual STATUS - *CLS The *CLS command clears all status data registers. Command Syntax *CLS Related Commands ALL_STATUS, CMR, DDR, *ES
WaveAce Remote Control Remote Control Through USB The USB interface is standard on every WaveAce oscilloscope. The rear panel USB port is the remote
WaveAce Remote Control STATUS - *ESR? The *ESR? query reads and clears the contents of the Event Status Register (ESR). The response represents the s
Operator's Manual Bit Value Name Description Note 3 8 DDE 1 - Device Dependent (specific) Error occurred. The Device specific Error (DDE) is set
WaveAce Remote Control STATUS - ALL_STATUS?, ALST? The ALL_STATUS? query reads and clears the contents of all status registers: ESR, INR, DDR, CMR,
Operator's Manual STATUS - CMR? The CMR? query reads and clears the contents of the CoMmand error Register (refer to the following table for deta
WaveAce Remote Control STATUS - DDR? The DDR? query reads and clears the contents of the Device Dependent or device specific error Register (DDR). In
Operator's Manual STATUS - EXR? The EXR? query reads and clears the contents of the EXecution error Register (EXR). The EXR register specifies th
WaveAce Remote Control Value Description 35 Waveform data error. Invalid waveform data have been detected. 36 Panel setup error. An invalid panel set
Operator's Manual STATUS - INR? The INR? query reads and clears the contents of the INternal state change Register (INR). The INR register record
WaveAce Remote Control WAVEFORM TRANSFER - STORE, STO The STORE command stores the contents of the selected trace. Command Syntax STOre [<trace>
Operator's Manual NOTE: The format of the waveform data depends on the current settings specified by the last WAVEFORM_SETUP command. Command Syn
Operator's Manual Remote Control Through LAN WaveAce 2000 series oscilloscopes have a standard LAN connection port for remote control that util
WaveAce Remote Control WAVEFORM TRANSFER – WAVEFORM SETUP, WFSU The WAVEFORM_SETUP command specifies the amount of data in a waveform to be transmitt
Operator's Manual Appendix: Waveform Template This section contains the Waveform Template describing the contents of the Waveform Descriptor prod
WaveAce Remote Control ; as their descriptions below. ; ;BLOCKS : ; < 36> WAVE_DESCRIPTOR: long ; length in bytes of block WAVEDESC < 40>
Operator's Manual < 68> RES_ARRAY2: long < 72> RES_ARRAY3: long ; 2 expansion entries are reserved ; ; The following variables identi
WaveAce Remote Control ; before first good point ; FIRST_VALID_POINT = 0 ; for normal waveforms. ; <128> LAST_VALID_PNT: long ; index of last g
Operator's Manual ; of the WFSU remote command. ; <144> SUBARRAY_COUNT: long ; for Sequence, acquired segment count, ; between 0 and NOM_SU
WaveAce Remote Control <164> MAX_VALUE: float ; maximum allowed value. It corresponds ; to the upper edge of the grid. ; <168> MIN_VALUE:
Operator's Manual ; <292> HORIZ_UNCERTAINTY: float ; uncertainty from one acquisition to the ; next, of the horizontal offset in seconds ;
WaveAce Remote Control _2 interpolated _3 sparsed _4 autoscaled _5 no_result _6 rolling _7 cumulative endenum ; <320> RESERVED5: word ; expansi
Operator's Manual _9 1_ns/div _10 2_ns/div _11 5_ns/div _12 10_ns/div _13 20_ns/div _14 50_ns/div _15 100_ns/div _16 200_ns/div _17 500_ns/div _1
WaveAce Remote Control 2. From the WaveAce menu, choose Utility > IP Settings (on the third page of settings). 3. Enter an IP Address, Subnet M
WaveAce Remote Control _35 500_ms/div _36 1_s/div _37 2_s/div _38 5_s/div _39 10_s/div _40 20_s/div _41 50_s/div _42 100_s/div _43 200_s/div _44 500_
Operator's Manual <332> FIXED_VERT_GAIN: enum _0 1_uV/div _1 2_uV/div _2 5_uV/div _3 10_uV/div _4 20_uV/div _5 50_uV/div _6 100_uV/div _7 2
WaveAce Remote Control _25 200_V/div _26 500_V/div _27 1_kV/div endenum ; <334> BANDWIDTH_LIMIT: enum _0 off _1 on endenum ; <336> VERTIC
Operator's Manual ; ; Explanation of the descriptor block USERTEXT at most 160 bytes long. ; ; < 0> TEXT: text ; a list of ASCII characters
WaveAce Remote Control RISTIME: ARRAY ; ; Explanation of the random-interleaved-sampling (RIS) time array RISTIME. ; This optional time array is only
Operator's Manual ; DATA_ARRAY_2: ARRAY ; ; Explanation of the data array DATA_ARRAY_2. ; This is an optional secondary data array for special ty
WaveAce Remote Control < 0> MEASUREMENT: data ; the actual format of a data is ; given in the WAVEDESC descriptor ; by the COMM_TYPE variable.
Operator's Manual Oscilloscope's TMPL? Query Response This template is the oscilloscope's response to a TMPL? query:/00 000000 LECROY_
WaveAce Remote Control ; ; where ; ; <byte position> = position in bytes (decimal offset) of the variable, ; relative to the beginning of the b
Operator's Manual ; s exponent fraction ; where ; s = sign of the fraction ; exponent = 11 bit exponent e ; fraction = 52 bit fraction f ; and th
Operator's Manual Header path Commands or queries that apply to a subsection of the oscilloscope, such as a single input channel, must have their
WaveAce Remote Control ; remote command COMM_FORMAT. ; text arbitrary length text string ; (maximum 160) ; unit_definition a unit definition consists
Operator's Manual In a way not following the byte boundaries, bits are then segregated as follows: 31, 30, 29 . . . . 24, 23, 22, 21 . . . . 2, 1