F8 is a view of the Differential Signal after it has been filtered by the 4.5GHz low-pass Butterworth
filter. This is calculated by subtracting the negative input from the positive input. This signal is
used to create many of the traces described earlier.
Figure 10 - OOB Test Results
In the Measure section:
IdleNoise (P1) is the peak-to-peak measurement of F3 (idle differential signal). This is the measured
value for Tx Max Noise During Idle (5.1.1). This value must be less than 120mV(P-P) in order
to pass the test.
OffsetDelta (P3) is the mean amplitude of F4 (burst differential signal). This is the measured value
for Tx- OOB Offset Delta (5.1.3). This value must be between -25mV and 25mV (inclusive) in
order to pass the test.
BurstAmpl (P4) is the peak-to-peak value of F4 (burst differential signal). This is the measured
value for Tx Max Burst Amplitude and Tx Min Burst Amplitude (5.1.2). The maximum
value must be less than 1.6V and the minimum value must be great than 240mV in order to pass
this test.
CMMdelta (P7) is the difference between the mean of the Burst Common Mode Signal (F5) and the
Idle Common Mode Signal (F7). This is the measured value for Tx - OOB Common Mode Delta
(5.1.4). This value must be between +/- 50mV in order to pass this test.