Operator’s Manual AP034 Active Differential Probe

Strona 2

AP034 Active Differential Probe 4 922261-00 Rev A Standard Accessories Hard Case ÷10 Plug-on Attenuator ÷20 Plug-on Attenuator Plug-on AC Coupler Pro

Strona 3

Operator’s Manual 922261-00 Rev A 5 Specifications Nominal Characteristics Nominal characteristics describe parameters and attributes that have are g

Strona 4 - Warranty

AP034 Active Differential Probe 6 922261-00 Rev A probe tip depends on proximity to probe cable ground, and is therefore non-reproducible. Teledyne L

Strona 5 - Table of Contents

Operator’s Manual 922261-00 Rev A 7 Figure 1, Typical CMRR Graphs General Characteristics Temperature: 0° to 50° C Operating -40° to 75° C St

Strona 6

AP034 Active Differential Probe 8 922261-00 Rev A Operation CAUTION: The input circuits in the AP034 incorporate components that protect the probe f

Strona 7 - Safety Instructions

Operator’s Manual 922261-00 Rev A 9 High common mode rejection requires precise matching of the relative gain or attenuation in the + and – input sig

Strona 8 - Operating Environment

AP034 Active Differential Probe 10 922261-00 Rev A Probe Input Loading Attaching any probe to a test circuit will add some loading. In most applicati

Strona 9 - Overview

Operator’s Manual 922261-00 Rev A 11 NOTE The AP034 Active Differential Probe transmits the measured signal differentially through the probe cable. T

Strona 10 - Optional Accessories

AP034 Active Differential Probe 12 922261-00 Rev A When using a differential probe or amplifier, be careful to avoid exceeding the common mode range.

Strona 11 - Specifications

Operator’s Manual 922261-00 Rev A 13 The AP034 offset allows you to measure signals up to 2.0 V (1.6 V of offset plus 400 mV of output signal). With

Strona 13 - Operator’s Manual

AP034 Active Differential Probe 14 922261-00 Rev A Adding Offset When the AP034 Active Differential Probe is used with a Teledyne LeCroy oscilloscope

Strona 14 - Operation

Operator’s Manual 922261-00 Rev A 15 NOTE: Probe offset is controlled with the channel OFFSET knob in oscilloscope firmware versions 8.1.0 or higher

Strona 15

AP034 Active Differential Probe 16 922261-00 Rev A Figure 5, Layout Dimensions for Test Fixtures Maintenance This probe is a high quality, precision

Strona 16 - Grounding the Probe

Operator’s Manual 922261-00 Rev A 17 service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory re

Strona 17 - Selecting the Proper Range

AP034 Active Differential Probe 18 922261-00 Rev A B. Incorrect Frequency Response Possible causes are a defective probe or oscilloscope, poor connec

Strona 18 - 12 922261-00 Rev A

Operator’s Manual 922261-00 Rev A 19 Returning a Probe Return a probe for calibration or service by contacting your local Teledyne LeCroy sales repre

Strona 19

AP034 Active Differential Probe 20 922261-00 Rev A 8. If returning a probe to a different country, also:  Mark shipments returned for service as a

Strona 20 - Adding Offset

Operator’s Manual 922261-00 Rev A 21 Table 3, Replaceable Part Numbers Reference Teledyne LeCroy Part Number Description Rev Qty. 1 529-304-001 Plas

Strona 21 - Autobalance

AP034 Active Differential Probe 22 922261-00 Rev A Matching Procedure for ÷10 Plugs The ÷10 and ÷20 Plug-on attenuators provided as standard accesso

Strona 22 - Maintenance

Operator’s Manual 922261-00 Rev A 23 Note: You can fabricate the Adjustment Tool by flattening the end of a 0.025 in. (0.635 mm) square pin with a f

Strona 23 - Troubleshooting

AP034 Active Differential Probe Operator’s Manual May 2013

Strona 24 - C. DC Errors

AP034 Active Differential Probe 24 922261-00 Rev A 11. Using the square adjustment tool, adjust the +LF Comp (See Figure 7) to achieve the best squar

Strona 25 - Returning a Probe

Operator’s Manual 922261-00 Rev A 25 Performance Verification This procedure can be used to verify the warranted characteristics of the AP034 Active

Strona 26 - Replacement Parts

AP034 Active Differential Probe 26 922261-00 Rev A Table 4, Performance Verification Test Equipment Description Minimum Requirements Test Equipment E

Strona 27

Operator’s Manual 922261-00 Rev A 27 Preliminary Procedure 1. Connect the AP034 Active Differential Probe to the female end of the ProBus Extension

Strona 28 - Equipment Required

AP034 Active Differential Probe 28 922261-00 Rev A 4. Connect one end of a second BNC cable to the output of the Function Generator. Attach the BNC-

Strona 29 - Procedure

Operator’s Manual 922261-00 Rev A 29 B. Check High Frequency Common Mode Rejection Ratio (CMRR) NOTE: Common Mode Rejection Ratio (CMRR) is defined a

Strona 30 - 24 922261-00 Rev A

AP034 Active Differential Probe 30 922261-00 Rev A 10. Move the high frequency sine wave generator output cable from the oscilloscope channel 2 input

Strona 31 - Performance Verification

Operator’s Manual 922261-00 Rev A 31 second BNC cable from the remaining end of the BNC Tee adapter to the Common Mode Drive No Termination input con

Strona 32 - 26 922261-00 Rev A

AP034 Active Differential Probe 32 922261-00 Rev A 15. Calculate the “Common Mode Rejection Ratio (CMRR) at 1 MHz” by dividing 8000 mV by the measure

Strona 33 - Preliminary Procedure

Operator’s Manual 922261-00 Rev A 33 Model: AP034 Serial Number: Asset or Tracking Number: Date: Technician: MODEL SERIAL NUMBER CAL

Strona 34 - 28 922261-00 Rev A

© 2013 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sale

Strona 35

AP034 Active Differential Probe 34 922261-00 Rev A Adjustment Procedure Introduction You can use this procedure to adjust the AP034 Active Differenti

Strona 36 - C. Check Low Frequency CMRR

Operator’s Manual 922261-00 Rev A 35 Test Equipment Required The table on the next page lists the test equipment and accessories, or their equivalent

Strona 37

AP034 Active Differential Probe 36 922261-00 Rev A Description Minimum Requirements Example Test Equipment Termination, precision 50 Ω ± 0.05% Teled

Strona 38 - 32 922261-00 Rev A

Operator’s Manual 922261-00 Rev A 37 Preliminary Procedure 1. Using pliers, carefully remove each of the push buttons from the ProBus interface hou

Strona 39

AP034 Active Differential Probe 38 922261-00 Rev A Procedure A. Adjust Coarse DC Balance (R226) 1. If necessary, set the probe offset to 0.000 V by

Strona 40 - Adjustment Procedure

Operator’s Manual 922261-00 Rev A 39 B. Adjust Fine DC Balance (R304) 1. Remove the shorting plug from the Logic Board. 2. Reinsert the shorting pl

Strona 41 - Test Equipment Required

AP034 Active Differential Probe 40 922261-00 Rev A C. Adjust Low Frequency CMRR (R7) 1. Leave the BNC cable connected to the probe end of the ProBus

Strona 42 - 36 922261-00 Rev A

Operator’s Manual 922261-00 Rev A 41 10. Set the oscilloscope to display channel 3. The waveform is the common mode signal. Turn the offset as necess

Strona 43

AP034 Active Differential Probe 42 922261-00 Rev A 5. Connect the other end of the cable to the DMM input, making sure that the plug corresponding t

Strona 44

Operator’s Manual 922261-00 Rev A 43 21. The trace should be offset from the center line by the same amplitude but opposite polarity of the value not

Strona 45

Operator’s Manual 922261-00 Rev A i Table of Contents Safety Instructions ...

Strona 46 - 40 922261-00 Rev A

AP034 Active Differential Probe 44 922261-00 Rev A F. Adjust Final Attenuator Compensation (C17) NOTE: The calibrator signal from a Teledyne LeCroy o

Strona 47 - D. Adjust OFFSET (R330)

Operator’s Manual 922261-00 Rev A 45 3. Set the oscilloscope display to channel 4, coupling to DC1MΩ , vertical scale factor to 200 mV/div, and the

Strona 48 - 42 922261-00 Rev A

AP034 Active Differential Probe 46 922261-00 Rev A I. Attenuator Matching and Final Check 1. Repeat the Performance Verification procedure to ensure

Strona 49 - E. Adjust GAIN (R322C)

Operator’s Manual 922261-00 Rev A 47 Reference Information Differential Mode and Common Mode Differential probes amplify the voltage difference that

Strona 50 - 44 922261-00 Rev A

AP034 Active Differential Probe 48 922261-00 Rev A Common Mode Rejection Ratio The ideal differential probe or differential amplifier would amplify o

Strona 51

Operator’s Manual 922261-00 Rev A 49 Certifications This section certifies the probe’s Electromagnetic Compatibility (EMC), Safety and Environmental

Strona 52 - 46 922261-00 Rev A

AP034 Active Differential Probe 50 922261-00 Rev A AUSTRALIA & NEW ZEALAND DECLARATION OF CONFORMITY—EMC The probe complies with the EMC provisio

Strona 53 - Reference Information

Operator’s Manual 922261-00 Rev A 51 Environmental Compliance END-OF-LIFE HANDLING The probe is marked with this symbol to indicate that it complies

Strona 54 - Common Mode Rejection Ratio

AP034 Active Differential Probe 52 922261-00 Rev A Contact Teledyne LeCroy Teledyne LeCroy Service Centers United States and Canada - World Wide Cor

Strona 56 - Safety Compliance

AP034 Active Differential Probe ii 922261-00 Rev A Matching Procedure for ÷10 Plugs ...

Strona 58 - Contact Teledyne LeCroy

Operator’s Manual 922261-00 Rev A 1 Safety Instructions This section contains instructions that must be observed to keep this oscilloscope accessory

Strona 59

AP034 Active Differential Probe 2 922261-00 Rev A Do not overload. To avoid electric shock, do not apply any potential that exceeds the maximum ratin

Strona 60

Operator’s Manual 922261-00 Rev A 3 Overview Description The AP034 is a wide band differential active probe. The probe features low noise, low input